PCount - Front Panel Parts Counting:

Front panel parts counting is a standard feature of the High Capacity Series balances. However, if an optional specific gravity

determination function (SHCSG) is purchased it is installed in place of the front panel parts counting function. High Capacity Series

balances can have either front panel parts counting or specific gravity determination, but not both.

User Action

Balance Response

1 Remove any weighing samples or containers from

the weighing pan, then press the TARE button.

Zeros are displayed.

2 Place the empty container in the center of the

weighing pan, then press the TARE button.

Zeros are displayed.

3 Press the MENU button repeatedly until PCount is

displayed.

PCount is displayed.

4

Press the PRINT button.

nc 10 is displayed. This indicates the number of samples required.

5 Press the MENU button repeatedly to change the

sample size.

The options of nc 10, nc 25, nc 50, nc 100 and MAnuAL are displayed.

6 Press the PRINT button to select the desired sample

size.

A series of dashes followed by LoAd and the required number of
samples are displayed.

7

Place the required number of samples in the container
and press the PRINT button.

LoAd will flash while the balance is processing the weight, then the
number of pieces is displayed.

8

Load the parts to be counted into the container.

The balance displays the number of parts.

9 If the weight of the individual samples is too light

for the balance.

Error is displayed. These parts must be counted on a balance with more
resolution.

10 If the individual sample weight is acceptable, but

the sample size is too small.

Add is displayed.

11 Place additional samples in the container and press

the PRINT button.

If the sample size is large enough the number of pieces is displayed. If
not, Add will be displayed.

12 Repeat step 11 until the sample size is large enough. The number of pieces is displayed.

13 To update the sample size at any time, place any

quantity, up to 255 pieces, into the container and
wait approximately 10 seconds.

The balance will display the number of samples for approximately 10
seconds, then display a series of ▼’s and will beep.

The sample size has now been updated to the displayed quantity.

14 To see the actual weight of the samples in the

container, press the MENU button.

The weight of the samples and a series of ▼’s is displayed.

15 To return to parts counting, press the MENU button

The number of pieces is displayed.

16 To exit the parts counting function, press the

ON/OFF button

The balance returns to the normal weighing mode.


UnitS - Unit of Weight Selection:

To change the units of measure proceed as follows.

• Repeatedly press the MENU button until UnitS is displayed.

• Press the PRINT button to select this function.

• Press the MENU button repeatedly to scroll through the available units of measure.

o GrAM

Grams Designated

by

a▼ on the display

o CArAt

Carats Designated

by

a▼ on the display

o OUncE

Ounces Designated

by

a▼ on the display

o Pound

Pounds Designated

by

a▼ on the display

o Penn

Pennyweights

o OUncEtr.

Troy ounces

o GrAin

Grains

o tAELHon

Hong Kong taels

o tAELSGP

Singapore taels

o tAELroc

R.O.C. taels

o MoMME

Momme

• Press the PRINT button to select the desired unit of weight and return the balance to its normal weighing mode.