Arge, Perture, Alorimeter – Scientech Large Aperture Calorimeter User Manual

Page 3: Pecifications, Ertification, Nvironmental, Equirements

Advertising
background image

3

L

ARGE

A

PERTURE

C

ALORIMETER

S

PECIFICATIONS

:

Model

360401

380401

380402

384UV5

360801

380801

380802

384UV5

Type of Absorber

Surface

Volume

Volume

Volume

Surface

Volume*

Volume

Volume

Aperture Diameter

100mm

200mm

Minimum Beam Diameter
Spectral Response

.25 - 35µm .266 - 1.2µm

9 -11µm

.193 - .36µm .25 - 35µm .266 - 1.2µm

9 -11µm

.193 - .36µm

Maximum Average Power

50W with full illumination of absorbing surface

100W with full illumination of absorbing surface

Minimum Average Power

150mW

700mW

Noise Level

1.5mJ - mW

7mJ - mW

Maximum Power Density

200W/cm²

See Note 1

4W/cm²

Note 2

200W/cm²

See Note 3

4W/cm²

Note 2

Maximum Peak Power Density

1MW/cm²

See Note 4 100MW/cm² See Note 5

1MW/cm²

See Note 6 100MW/cm² See Note 5

Maximum Single Pulse Energy
Maximum Energy Density

Note 7

Note 8

4J/cm

2

Note 9

Note 7

Note 10

4J/cm

2

Note 9

Precision

< 1%

Accuracy

5%

Response Time

5 sec when connected to a Scientech Indicator in Watts Mode

Dimensions DxL - inches/cm

6.00 x 8.00/15.24 x 20.32

9.00 x 10.00/22.86 x 25.40

Weight - pounds/kgs

6/2.72

16.27/7.26

Indicator Compatibility

H410, H410D, S310, S310D

* This is a segmented absorber
Note 1: 380401

27W/cm

2

@ 1064 nm, 21W/cm

2

@ 532 nm, 7.7W/cm

2

@ 355 nm, 158mW/cm

2

@ 266nm

Note 2: 384UV5, 388UV5

50W/cm

2

@ 355nm

Note 3: 380801

13.5W/cm

2

@ 1064 nm, 10.5W/cm

2

@ 532 nm, 3.85W/cm

2

@ 355 nm, 79mW/cm

2

@ 266nm

Note 4: 380401

90GW/cm

2

@ 1064 nm, 71GW/cm

2

@ 532 nm, 27GW/cm

2

@ 355 nm, 530MW/cm

2

@ 266nm

Note 5: 384UV5, 388UV5

Repetitive pulses: 101MW/cm

2

@ 355nm

Single pulses: 3.5GW/cm

2

@ 355nm

Note 6: 380801

45GW/cm

2

@ 1064 nm, 35.5GW/cm

2

@ 532 nm, 13.5GW/cm

2

@ 355 nm, 265MW/cm

2

@ 266nm

Note 7: 360401, 360801

Max J/cm

2

= 1000 x (pulse width)

1/2

to a maximum of 200J/cm

2

Note 8: 380401

Repetitive pulses:
3.7J/cm

2

@ 1064nm, 2.9J/cm

2

@ 532nm, 1J/cm

2

@ 355nm, 20mJ/cm

2

@ 266nm

Single pulses:
7J/cm

2

@ 1064nm, 5.6J/cm

2

@ 532nm, 2.1J/cm

2

@ 355nm, 41mJ/cm

2

@ 266nm

Note 9: 384UV5, 388UV5

Repetitive pulses: 1.1J/cm

2

@ 355nm

Single pulses: 40J/cm2 @ 355nm

Note 10: 38-0801

Repetitive pulses:
1.85J/cm

2

@ 1064nm, 1.45J/cm

2

@ 532nm, 0.5J/cm

2

@ 355nm, 10mJ/cm

2

@ 266nm

Single pulses:
3.5J/cm

2

@ 1064nm, 2.8J/cm

2

@ 532nm, 1.05J/cm

2

@ 355nm, 20.5mJ/cm

2

@ 266nm

150J

300J

5cm

7.5cm

CE

M

ARK

C

ERTIFICATION

:

All of the calorimeters listed in this manual have been certified for the European CE mark.


E

NVIRONMENTAL

R

EQUIREMENTS

:

This product is intended for indoor use at altitudes up to 2000 meters, Pollution Degree 2 in accordance with
IEC 664 and transient overvoltages according to Installation Categories (Overvoltage Categories) II. Note that
each of the above detectors will not pass the IEC 801 Publication, Part 3, Radiated Electromagnetic Field
Requirements. The system, meter and detector, is designed to measure radiation within the test's radiation band.
The detectors were held outside the radiated electromagnetic field during this test. It is up to the user to be
aware of RF fields present during measurements and their effects if any on those measurements.