F. Record the voltage reading of the DVM (V

c

).

G. Calculate the calorimeter’s output sensitivity (S) as follows:

S = V

c

/W

where:

S = calorimeter’s output sensitivity

V

c

= voltage output from the calorimeter in mV

W = desired laser power output.


The measured sensitivity should be ± 3 % of the calorimeters original sensitivity value.

F

ACTORY

R

ECALIBRATION

:

Scientech recommends that a complete calibration be performed annually to verify system accuracy. Please
contact our Product Service Department at (800)525-0522 or (303)444-1361 or Fax (303)444-9229 or email
[email protected] to arrange for a NIST traceable, factory calibration.

L

IMITED

W

ARRANTY

:

All Scientech Laser Power and Energy Measurement Systems are warranted against defects in materials and
workmanship for two (2) years from date of delivery. During the warranty period, Scientech will repair, or at its
option replace at no charge, components that prove to be defective. The equipment must be returned, shipping
prepaid, to Scientech's product service facility. This limited warranty does not apply if the equipment is
damaged by accident or misuse or as a result of service or modification by other than a Scientech service
facility. The foregoing warranty is in lieu of all other warranties expressed or implied including but not limited
to any implied warranty of merchantability, fitness, or adequacy for any special incidental or consequential
damages whether in contract, tort, or otherwise.

R

ETURNED

G

OODS

P

ROCEDURE

:

Should it become necessary to return any item to Scientech for any reason, please contact our Product Service
Department at (800)525-0522 or (303)444-1361 or Fax (303)444-9229 or email [email protected]. When
you call, please be ready to provide model number, serial number, and a description of the problem. Frequently
we can provide self-help information which will eliminate the need for returning the unit(s).

If equipment return is required, please pack the items in the original box and packing material. As an alternate,
place the equipment in a snug-fitting box, and then pack that box in a larger box with at least four inches of
packing material. Scientech does not assume responsibility for under packed items.

Please include the name and phone number of the person we should contact regarding repair questions.

Normally, products are repaired and shipped within 5 working days after their arrival at the product service
facility. This is an average time and could vary depending on the workload.