Page 33: Serial interface parameters, 11 serial interface parameters

Advertising
background image

1225

Flashes the current value and then displays the value for
adjustment

1225.0

Current Soft Cutout setting

Adjust this parameter by using “UP”, “DOWN”, and “SET” as each digit
flashes.

1200.0

New Soft Cutout setting

Press “SET” to accept the new temperature limit.

If the temperature of the unit is ever greater than the Soft Cutout temperature
the controller shuts itself down and displays, alternately, “SCtOut” and “Err
8”.

8.10.3

Cutout Reset Mode

The cutout reset mode determines whether the cutout resets automatically when
the well temperature drops to a safe value or must be manually reset by the
operator.

The parameter is indicated by,

CtorSt

Cutout reset mode parameter

Press “SET” to access the parameter setting. Normally the cutout is set for au-
tomatic mode.

Auto

Cutout set for automatic reset

To change to manual reset mode press “UP” or “DOWN” and then “SET”.

rSt

Cutout set for manual reset

8.11

Serial Interface Parameters

The serial RS-232 interface parameters menu is indicated by,

SEriAL

Serial RS-232 interface parameters menu

The serial interface parameters menu contains parameters which determine the
operation of the serial interface. The parameters in the menu are: BAUD rate,
sample period, duplex mode, and linefeed.

9150 Portable Furnace
User’s Guide

32