Issue

Recommendations

Recharge the battery. Plug the indicator into 110 VAC for 18+
hours (using the included AC adapter). When not operating under
battery power, it is recommended that the indicator remains
plugged in to 110 VAC. This will keep a full charge in the battery.

Indicator displays a weight after the animal
has left the scale.

Check for mechanical binding in the overall equipment set up,
such as restraining chains on squeeze chute, platform rubbing
against walls, or similar.

Check weighing platform area for build up of mud or manure.

Press the Zero key to zero out the scale.

Weight reads out lower at one end of the
weighing platform than the other end.

Check for any type of mechanical binding or impingement of the
weigh bar that is displaying the lower weight. Check underneath
the weigh bar for any obstructions. Make sure that the weigh bar
feet are not screwed in so far as to restrict downward movement
of the weigh bar.

Inspect the weigh bar cables for damage, such as flat spot(s),
split outer sheath, sharp kinks, looseness of cable connections,
etc. If damaged, the cable must be replaced.

If the problem persists, contact Triner Scale Tech Support.

Indicator displays six small “u” characters.

Scale is overloaded. Remove weight from scale.

Indicator displays six small “n” characters.

Scale is underloaded. Power off, then power on the scale.

Displayed weight continually drifts.

Inspect the weigh bar cables for damage, such as flat spot(s),
split outer sheath, sharp kinks, looseness at weigh bar connection,
etc. If damaged, the cable must be replaced.

If the problem persists, contact Triner Scale Tech Support.

Scale turns off on its own.

The indicator has a power conservation feature, set to automatically
power off the scale after 10 minutes of non-use. If your needs
require a different setting, call Triner Tech Support.

Display is erratic.

The battery may be fully depleted. If so, this condition can cause
erratic displays even if the digital indicator is plugged into 110
VAC.

Power off the indicator and fully charge the battery for 18+ hours.

Inspect the weigh bar cables for damage, such as flat spot(s),
split outer sheath, sharp kinks, looseness at weigh bar connection,
etc. If damaged, the cable must be replaced.

If the problem persists, contact Triner Scale Tech Support.

Triner Scale Toll-free Tech Support:

(800) 238-0152

Low battery icon (

) flashes.

TROUBLE SHOOTING

Alley-Bar™
Livestock Weighing System