Calibration in the factory – VACUUBRAND PC 2001 VARIO User Manual

Page 45

Advertising
background image

page 45 of 49

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 999100 / 06/05/2008

Cleaning the gauge head VSK 5

The gauge itself is maintenance-free.

If the vacuum system is contaminated (oil, particles, etc.), contamination of the pressure transducer will
influence the accuracy of measurement.

➨ Fill the gauge head chamber with a solvent (e. g. benzene) and allow sufficient cleaning time. Observe

all regulations concerning usage and disposal of solvents!

➨ Drain the solvent and dispose of in accordance with regulations, repeat cleaning if necessary.

➨ Rinse the gauge head chamber several times with alcohol in order to remove all solvent residues.

☞ Never use a spiky or sharp-edged tool to clean the pressure transducer.

➨ Allow the pressure transducer to dry.

➨ Readjust the pressure transducer if necessary.

Calibration in the factory

Control of measuring equipment

The VACUUBRAND DKD calibration laboratory is accredited by the Physikalisch-Technische Bundesanstalt
(PTB; German national institute for science and technology and the highest technical authority of the
Federal Republic of Germany for the field of meteorology and certain sectors of safety engineering) for the
measurable variable pressure in the pressure range from 10

-3

mbar to 1000 mbar in accordance with

the general criteria for the operation of testing laboratories defined in the DIN EN ISO/IEC 17025:2000
series of standards.
Calibration in the VACUUBRAND calibration laboratory:
-

To meet the requirements of the DIN ISO 9000ff and 10012 series of standards regarding the calibration
of inspection, measuring and test equipment at specified intervals.

-

To document that the vacuum gauges calibrated are traceable to national standards of the PTB.