VACUUBRAND PC 2001 VARIO User Manual

Page 6

Advertising
background image

page 6 of 49

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 999100 / 06/05/2008

In case of blockade of the motor (after 10 attempts to start-up) the pump is switched off.
☞ If the pump is switched off due to safety measures, manual reset is necessary.

Isolate the pump from mains. Eliminate cause of failure before restarting the pump.

☞ Avoid high heat supply (e. g. due to hot process gases).

☞ Ensure sufficient air admittance if pump is installed in a housing.

Due to the residual leak rate of the equipment, there may be an exchange of gas,
albeit extremely slight, between the environment and the vacuum system.
☞ Adopt suitable measures to prevent contamination of the pumped substances or

the environment.

Pumping at high inlet pressure may lead to overpressure at the gas ballast valve.
☞ Pumped gases or condensate might be pushed out if the valve is open.

☞ If an inert gas supply is connected, ensure that the inlet pipeline is not contaminated.

The controller is equipped with a short circuit proof transformer with an integrated
overload protection (no fuses).
☞ Failure of the pump (e. g. by power failure) or connected components, parts of the

supply (e. g. coolant) or change of parameters (e. g. increase of pressure in the coolant
system) must not lead to a critical dangerous situation under any circumstances.

Electronic equipment is never 100% fail-safe. This may lead to an indefinite status of
the equipment. Provide protective measures against misfunction and failure.
☞ Operating the pump with high or low frequency, stand still of the pump or operating

the venting valve must not lead to a critical dangerous situation under any circum-
stances.

Ensure that in case of failure the pump and the vacuum system always will turn into a
safe status.
☞ In case of diaphragm cracks or leaks in the manifold pumped substances might be

released into the environment or into the pump housing. To reduce the risk of leaks,
ask for a diaphragm pump with additional safety diaphragm.

☞ Comply especially with notes on operation and use and maintenance.

Use only genuine spare parts and accessories.
☞ Otherwise safety and performance of the equipment as well as the electromagnetic

compatibility of the equipment might be reduced.

☞ Possibly the CE mark or the C/US conformity becomes invalid if not using genuine

spare parts.

Ensure that maintenance is done only by suitably trained and supervised technicians.
Ensure that the maintenance technician is familiar with the safety procedures which
relate to the product processed by the vacuum system and that the equipment, if
necessary, is appropriately decontaminated before starting maintenance.
Comply with local and national safety regulations.

Wear parts have to be replaced regularly. In case of normal wear the lifetime of the
diaphragms and valves is > 10000 operating hours. Bearings have a typical durability
of 40000 h.
Before starting maintenance vent the system, isolate the pump and other components
from the vacuum system and the electrical supply, drain condensate and allow suffi-
cient cooling of the pump.
Before starting maintenance, wait two minutes after isolating the equipment from mains
to allow the capacitors to discharge.
Never operate the pump if covers or other parts of the pump are disassembled.
Never operate a defective or damaged pump.
Repair of the gauge head VSK 5 is not possible.
In order to comply with law (occupational, health and safety regulations, safety at work
law and regulations for environmental protection) vacuum pumps, components and
measuring instruments returned to the manufacturer can be repaired only when certain
procedures (see section “Notes on return to the factory“) are followed.

Advertising