If pumping condensable vapors (water vapor,

solvents, etc.), let the pump run with

gas bal-

last to help purge any condensation in the

pump.

gas ballast

(open)

Prevent internal condensation, transfer of liq-

uids or dust. The diaphragms and valves will

be damaged, if liquids are pumped in signifi-

cant amounts.

Check the pump regularly for external soiling

and deposits. Clean the pump if necessary

to avoid an increase of the pump’s operating

temperature.

NOTICE

IN

condensate / dust

In case of overload, the motor is shut down by

a

self-hold thermal cutout in the winding.

Note: Only manual reset is possible. Switch

off the pump and disconnect from the power

source. Identify and eliminate the cause of fail-

ure. Allow the pump to cool down sufficiently

before restart.

°C / °F

I ➨ O

!

3.

2.

1.

IN

Avoid overheating (e.g., due to hot process

gases). Maximum permissible temperature

range: see “Gas inlet temperatures”, pg. 27.

Make sure the fans’ air supply is adequate.

Do not place the pump on soft surfaces (e.g.,

rubber foam) during operation. This may re-

duce the air circulation in the pump housing

and affect the fans’ cooling effect. Check fans

regularly for dust/dirt. Clean fan guard grills if

necessary to avoid a reduction of ventilation.

air

air

NOTICE

NOTICE

NOTICE