Troubleshooting – VACUUBRAND MV 10C NT VARIO o.C. User Manual

Page 48

Advertising
background image

page 48 of 97

Troubleshooting

Fault

Possible cause

Remedy

❑ Pump does not

start or stops im-

mediately.

➨ Electrical power cord

not plugged in, electrical

supply failure?

✔ Plug in power cord.

Check fuse.

➨ Device fuse blown?

✔ Identify cause of failure.

Replace device fuse.

➨ Overpressure in outlet

line?

✔ Remove blockage in line,

open valve.

➨ Motor overloaded?

✔ Allow motor to cool down,

identify and eliminate

cause of failure. Manual

reset is necessary. Switch

off pump or unplug.

❑ Pump does not

achieve its ulti-

mate vacuum or

usual pumping

speed.

➨ Centring ring at small

flange connection not

correctly positioned, or

leak in the pipeline or

vacuum system?

✔ Check pump directly -

connect vacuum gauge

directly at pump inlet -

then check connection,

pipeline and vacuum

system if necessary.

➨ Long, narrow vacuum

line?

✔ Use lines with larger di-

ameter, length as short as

possible.

➨ Pump has been exposed

to condensate?

✔ Allow pump to run for

some minutes with atmo-

spheric pressure at the

inlet to purge.

➨ Deposits have been

formed inside the pump?

✔ Clean and inspect the

pump heads.

➨ Diaphragms or valves

damaged?

✔ Replace diaphragms and/

or valves.

➨ Outgassing substances

or vapor generated in

the process?

✔ Check process parame-

ters.

➨ Pump too hot?

✔ Allow pump to cool down.

Determine and eliminate

the cause of overheating.

❑ Pump too noisy.

➨ Atmospheric or high

pressure at the pump

inlet?

✔ Connect hose or silencer

to pump outlet. Be careful

not to cause outlet over-

pressure, especially with

condensable vapors.

Advertising