Maintenance and repair, Notice – VACUUBRAND MZ 2 VARIO-SP User Manual

Page 18

Advertising
background image

page 18 of 51

Adopt suitable measures to prevent any danger aris-

ing from hot surfaces or electric sparks. Ensure that

hot surfaces of the pump do not cause burns. Provide

a suitable contact guard if necessary.

Prevent the backpressure of gases and the backflow of

condensates at the outlet.

Provide appropriate protective measures to allow for the

possibility of failure and

malfunction. The protective mea-

sures must also allow for the requirements of the respec-

tive application.
A temperature sensor at the circuit board protects the mo-

tor: Current limitation in case the temperature at the circuit

board raises above 158°F (70°C). At very high tempera-

tures the pump is switched off.

Note: Only manual reset is possible. Disconnect the pump

from the power source. Identify and eliminate the cause of

failure.

Maintenance and repair

In order to comply with laws (occupational, health and

safety regulations, safety at work law and regulations for

environmental protection) vacuum pumps, components

and measuring instruments can only be returned when

certain procedures (see section „Replacing diaphragms

and valves“, pg. 38) are followed.

Take advantage of our service seminars, which put special

focus on the maintenance and repair of vacuum pumps.

For details and for the online ”Instructions for repair” man-

ual see www.vacuubrand.com.

In normal use, the lifetime of the diaphragms and valves

is typically 15,000 operating hours (at 1500 rpm). Bear-

ings have a typical durability of 40000 h.