Controlling the pump

Connecting the cable:

Use and operation

As set at the factory, the pump is designed for operation

with a control signal; i.e., without a control signal, the pump

does not start!

To operate the pump without a control signal, change the

settings on the circuit board (see below)!

External setting of the motor speed via PWM (factory set):

PWM (5V - max. 24V), frame frequency: 100Hz to 1.5kHz

0% to 100% PWM: Linear increase of the motor speed (0 to 2400 rpm):

Minimum motor speed 400rpm: at 17% PWM

Maximum motor speed 2400rpm: at 100% PWM

e.g., at a frame frequency of 100 Hz and 25% PWM: pulse-width 2.5 ms

NOTICE

The control line contains four differently colored wires.

wire in control line

assignment

red

+24V DC (supply voltage, max 7A)

blue

GND (24V)

white

voltage input: 0V to 10V DC or

PWM: 5V to max. 24V

(depending on the control signal input for

motor speed)

black

GND Signal

Selecting the control signal input:

(PWM / voltage / internal setting)
To select a control signal input other than the factory set one (PWM), the

control line at the terminal board has to be reconnected.