During operation:

Max. ambient temperature: 40 °C
☞ Make sure ventilation is adequate if pump is installed in a housing or if ambient

temperature is elevated.

☞ If pump is installed in altitudes of more than 1000 m above mean sea level check

compatibility with applicable safety requirements, e. g. IEC 60034 (motor may over-
heat due to insufficient cooling).

The motor is shut down by a thermal cutout in the winding.
☞ Attention: Manual reset is necessary. Switch off the pump or isolate the equipment

from mains. Wait approx. five minutes before restarting the pump.

Attention: In case of supply voltage below 100V, the lock of the cutout might be

restricted and the pump might restart on its own after sufficient cooling down. Take
suitable precautions, if an automatic restart of the pump may lead to a critical
dangerous situation.

Due to the high compression ratio of the pumps, pressure at the outlet port might be
generated being higher than the max. permitted pressure compatible with the mechani-
cal stability of the system.

Do not start pump if pressure difference between inlet and outlet port exceeds
max. 1 bar
. Attempts to start pump at higher difference may cause blockade and
damage of the motor.
☞ Check compatibility with max. permitted pressure at outlet and max. pressure

difference between inlet and outlet ports.

Prevent internal condensation, transfer of liquids or dust. The diaphragm and valves
will be damaged, if liquids are pumped in significant amounts.

A warm up period (approx. 15 min.) is required to ensure that the rated ultimate total
pressure and pumping speed are attained.

Shutdown:

Has the pump been exposed to condensate?
☞ Allow the pump to continue to run at atmospheric pressure for a few minutes.

Has the pump been exposed to media which may damage the pump materials or form
deposits?
☞ Check and clean pump heads if necessary.