Safety information, General information, Intended use – VACUUBRAND GKF 1000i User Manual

Page 3: Setting up and installing the cold trap gkf 1000i, Ambient conditions, Notice

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users’ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999076 / 19/09/2012

page 3 of 8

Safety information!

General information

Read this manual carefully before installing or operating the equipment.

Observe the instructions contained in this manual.

Comply with technical data and notes on use and operation.

Do not use any damaged equipment.

• Prior to every use: Check the cold trap for damage. The glass surfaces must be

free from damage, notches, cracks or scrapers.

Remove the product from its packing-box. If the equipment is damaged, notify the

supplier and the carrier in writing within three days; state the item number of the

product together with the order number and the supplier’s invoice number. Retain all

packing material for inspection. In case, mount enclosed components.

Intended use

Comply with notes on correct vacuum connections, see section ”Use and opera-

tion”.

Use the cold trap for the intended use only, i.e., for the condensation of vapours in

vacuum installations.

Suitable coolants: Liquid nitrogen (LN

2

) or acetone-dry ice (CO

2

) mixtures.

Setting up and installing the cold trap GKF 1000i

Prevent any part of the human body from coming into contact with vacuum or with

cryogenic coolant.

• Assemble the cold trap to a base plate or or fixate at a stand mounting. Provide a

firm, level platform. Check that the system to be evacuated as well as the hose and

ground joint connections are mechanically stable and that all fittings are secure.

Note: Flexible elements will shrink when evacuated.

• Prevent mechanical stress at the connection due to tensile force or vibrations.

• Connect hoses gas tight towards the vacuum installation and towards the inlet of

the vacuum pump.

• Maximum operating pressure:

1.1 bar absolute.

Comply with all applicable and relevant safety requirements (regulations and guide-

lines). Implement the required actions and adopt suitable safety measures.

Ambient conditions

Pay attention to the

permissible maximum ambient temperatures between

+10°C and +40°C (see ”Technical data”).

NOTICE

NOTICE