The drive and all bearings are encapsulated and are packed with life-long lubricant.
Under normal operating conditions, the pump is maintenance-free. The valves and
diaphragms are wear parts. Aging of the oil in the rotary vane pump necessitates oil
changes.

Note: Any drop in the pumping speed of the diaphragm pump produces an increase in
pressure in the oil reservoir. Although this does not have a direct effect on the pump-
ing speed and ultimate vacuum attainable by the HYBRID pump, it does have a con-
siderable effect on the aging of the oil and the HYBRID pump’s chemical resistance.

Shutdown:

Has the diaphragm pump been exposed to condensate or was the pump used to
pump vapours?
☞ Allow pump to run on with gas ballast valve initially open, and then with gas ballast

valve closed.

Short-term:
☞ Restart not possible until 2 minutes have elapsed (cool-down phase of start-up

controller).

Long-term:
➨ Drain old oil.

➨ Pour in a small quantity of new oil and briefly operate pump (to flush system).

➨ Drain oil.

➨ Fill pump with new oil.

Before transporting pump, drain condensate from catch pots of separator on inlet
side and oil mist filter from outlet side.

Important: When transporting a pump containing oil, take great care not to tilt the
pump to an angle that would permit oil to make its way into the filter element. Drain oil
if necessary.