Before starting maintenance, isolate the pump from the vacuum system and the elec-
trical supply so that the pump cannot be operated accidentally.

Has the pump been exposed to dangerous or corrosive gases?
Take appropriate safety measures (e. g. safety-clothing and protective goggles) to
avoid inhalation and skin contact.
The pump and the oil may be contaminated with pumped chemicals which have been
pumped, adopt suitable decontamination measures if necessary.

Aging of the oil necessitates oil change in case of darker colour (see new oil), strange
odour of the oil or foreign particles.
Depending on individual cases (especially if corrosive gases or vapours have been
pumped) it may be efficient to check the oil at appropriate intervals.

The oil contains small quantities of water/solvent:
☞ The oil can be cleaned to a certain extent by operating pump for 1 - 2 hours with

the inlet line sealed and gas ballast valve open.

☞ After every oil change (with the exception of the first oil change after 100 operating

hours), inspect and clean pump chambers of diaphragm pump.

☞ After every oil change, check condition of filter element in oil mist filter; replace

element if necessary.

Under normal operating conditions:
➨ Check oil level every time before starting the pump.

☞ Carry out the first oil change after 100 hours operation time.

☞ Change oil on a yearly basis at the latest.

☞ Dispose of the used oil, which may be contaminated by chemicals, according to all

applicable regulations.

Oil change

oil outlet

sight glass for oil level

oil inlet

rating plate with marks
Min./Max.
for oil level