Safety information!

General information

Read and comply with this manual before installing or operating the equip-

ment.

Comply with technical data and notes on use and operation.

If the equipment is damaged, notify the supplier and the carrier in writing within three

days; state the item number of the product together with the order number and the

supplier’s invoice number. Retain all packing material for inspection.

Do not use the equipment if it is damaged.

Intended use

Use the cold trap for the intended use only, i. e. for the condensation of vapours in

vacuum installations.

Suitable coolants: Liquid nitrogen (LN

2

) or acetone-dry ice (CO

2

) mixtures.

Setting up and installing the cold trap

Prevent any part of the human body from coming into contact with vacuum or with

cryogenic coolant.

• Connect pipes gas tight to the vacuum installation and to the inlet of the vacuum

pump.

• Prevent mechanical stress at the connection due to tensile force or vibrations.

Attention: Flexible elements tend to shrink when evacuated.

• Maximum operating pressure:

1.1 bar absolute.

Fixate the cold trap at a stand mounting.

Check that the system to be evacuated as well as the small flange connections are

mechanically stable and that all fittings are secure. Comply with all applicable safety

regulations.

Comply with notes on correct vacuum connections, see section ”Use and opera-

tion”.

Comply with all

applicable and relevant safety requirements (regulations and

guidelines),

implement the required actions and adopt suitable safety meas-

ures.

Ambient conditions

Pay attention to the

permissible maximum ambient temperatures between +10°C

and +40°C (see ”Technical data”).