Operating conditions of cold trap SKF H 25 / SKF H 40

The cold trap SKF H 25 / SKF H 40 is

not suitable to condensate

- unstable substances,

- self inflammable substances,

- substances which are inflammable without air and

- explosive substances.

Safety during operation

Adopt suitable measures to prevent the release of dangerous, toxic, explosive,

corrosive, noxious or polluting fluids when draining condensate.

Adopt suitable measures to prevent the formation of explosive or flammable mix-

tures. Oxygen may condensate at the temperature of liquid nitrogen, use inert

gas for venting if necessary.

A great amount of gas can be absorbed on cold surfaces. The gases may ex-

pand abruptly in case of warming. This may lead to inadmissible overpressure in

the system. Risk of bursting!

Check coolant level in the cold trap during operation at appropriate intervals.

• Check the cold trap for damage. Do not use damaged components.

• Use only

genuine spare parts and accessories.

Otherwise safety and performance of the equipment might be reduced.

• Comply with all applicable safety measures and requirements when using cryo-

genic coolants.

Use only transport receptacles intended for coolants.

Wear safety glasses and protective gloves.

• Do not clamp the covers of coolant receptacles. Ensure pressure compensation

between coolant receptacle and atmosphere.

• Comply with applicable regulations when disposing of condensates. Take into

consideration that the condensate may be polluted.

Take adequate precautions to protect yourself and other people from the effects

of dangerous substances (inhaling or skin contact). Wear appropriate safety-

clothing and safety glasses.

Check condensate level regularly and drain condensate in time (condensate drain

screw).

To the best of our knowledge the cold trap SKF H 25 / SKF H 40 is in compliance with

the requirements of the applicable standards and directives.