Vacuum management with vms module a – VACUUBRAND CVC 2II User Manual

Page 16

Advertising
background image

page 16 of 36

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 997835 / 15/12/2009

Attention: The basic setups Management, Management

Plus or VACUU•LAN are selectable only if a

VMS Module A and an isolation valve if necessary are connected!

Notes on choosing the basic setup

The controller CVC 2

II

can be adapted to the specific application by choosing the

appropriate basic setup Standard, Management, Management

Plus or VACUU•LAN

and choosing an operation mode, two-point control or continuous pumping: The con-
nected components of the chemistry vacuum system have to be preset once only.

The vacuum controller CVC 2

II

offers four basic setups depending on

-

the components of the VACUUBRAND chemistry vacuum system which are con-
nected to the system.

-

the specific user and/or process requirements.

The controller CVC 2

II

identifies automatically if an isolation valve is connected. If it is

missing, the basic setups “Standard“ and “Management

Plus“ are not selectable.

Warning triangle and symbol isolation valve are flashing:
➨ Isolation valve is missing, is however indispensable for the

selected basic setup.

☞ Connect isolation valve or choose appropriate basic setup.

Vacuum management with VMS Module A