Readjustment of the vacuum gauge dvr 2, Adjustment at atmospheric pressure, Adjustment under vacuum – VACUUBRAND DVR 2 User Manual

Page 11

Advertising
background image

page 11 of 17

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.:

999001 / 04/04/2012

Readjustment of the vacuum gauge DVR 2

Press ”UP/DOWN” key and keep it pressed

while switching on.

The vacuum gauge switches into the adjust-

ment mode (indicated by a warning triangle).

The arrow at the right indicates, that the pres-

The arrow at the right indicates, that the pres-

sure reading can be increased using ”UP/

DOWN” key.

Press ”UP/DOWN” key to increase the read-

ing to the actual atmospheric pressure.

The equipment has been adjusted at the factory. In general, there is no need for adjustment by the user

due to the excellent longtime stability.

Depending on operation conditions, type of application and accuracy requirements, an inspection and

readjustment may become necessary.

Admit air to the vacuum gauge. Make sure that the vacuum connection at the vacuum gauge is at at-

mospheric pressure.

Attention: Determine the exact actual atmospheric pressure, e. g. by using an accurate barometer or

get accurate reading from the weather service, e. g. at the next airport, etc. (take into account the differ-

ence in altitude between e. g. airport and laboratory).

To reduce the reading:

Press ”MODE” key to change the direction

of adjustment. The arrow moves to the left

indicating that the pressure reading can be

reduced.

Press ”UP/DOWN” key to reduce the reading.

Evacuate the vacuum gauge to a pressure <0.5 mbar (<0.4 Torr) e. g. by applying a good rotary vane

pump.

Press ”UP/DOWN” key and keep it pressed

while switching on.

The vacuum gauge switches into the adjust-

ment mode (indicated by a warning triangle).

The reading is adjusted automatically to

”zero”.

Attention: Adjustment under vacuum with a pressure higher than 0.5 mbar (0.4 Torr) reduces the ac-

curacy of measurement. If the pressure is significantly higher than 0.5 mbar (0.4 Torr), adjustment to a

reference pressure is recommended

.

Press ”ON/OFF” key to confirm the adjust-

ment and to terminate the adjustment mode.

Adjustment at atmospheric pressure

+

Press ”ON/OFF” key to confirm the adjust-

ment and to terminate the adjustment mode.

Adjustment under vacuum

+