page 13 of 17

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.:

999001 / 04/04/2012

Note: Some of the components on the printed circuit board are sensitive to electrostatic discharge.

Do not touch components. If necessary, ground the printed circuit board and the person replacing the

battery suitably.

Attention: Use only a battery of the recommended type (see ”Technical data”).

The battery must not be charged or connected to an external voltage.

Remove the battery.

Insert the new battery into the battery compartment cor-

rectly matching the positive and negative polarity (as

shown by the engraving inside the compartment).

When reassembling the vacuum gauge, make sure that

the switch pad and the printed circuit board are positioned

correctly.

Carefully screw both parts of the housing together.

battery

switch pad

printed circuit board

Replacing the battery

Unscrew the four countersunk head screws at the rear

side of the housing with a Phillips screw driver.

Disassemble the housing parts carefully.

Attention: The two housing parts are connected by the

sensor cable.

Advertising