Cleaning the gauge head

The gauge itself is maintenance-free.

Only VSP 5: Using small screw driver or pencil tip, release locking ring in gauge head connection and

remove protective filter with shim, keep locking ring.

➨ Fill the gauge head chamber with a solvent (e. g. benzene) and allow sufficient cleaning time. Observe

all regulations concerning usage and disposal of solvents!

➨ Drain the solvent and dispose of in accordance with regulations, repeat cleaning if necessary.

➨ Rinse the gauge head chamber several times with alcohol in order to remove all solvent residues.

➨ Allow the pressure transducer to dry.

Only VSP 5: Assemble protective filter and locking ring, replace protective filter if necessary.

➨ Readjust gauge head if necessary.

Note:
The gauge head VSP 5 can be used without protective filter, if only the protective filter is not compatible
with the pumped media (see “Technical data“).

If the vacuum system is contaminated (oil, particles etc.):
☞ Replace the protective filter (cat. no.: 63 87 29) in the gauge head VSP 5 if necessary.

➨ Using small screw driver or pencil tip, release locking ring in gauge head connection and remove

protective filter with shim, keep locking ring.

Readjustment

Readjusting the gauge head VSP 5

Readjustment at atmospheric pressure can be performed easily. Readjustment “vacuum“ at a pressure
< 5 x 10

-4

mbar should only be performed, if that vacuum range can be reached with a vacuum pump

available. Readjustment under vacuum is only necessary when, due to aging or contamination of the
filament, incorrect readings are observed. Clean a contaminated gauge head prior to readjustment.

Adjustment at atmospheric pressure:

➨ Vent vacuum system or gauge head, make sure, that gauge head is exposed to atmospheric pressure.

➨ Push key-button at the side of the housing with a pencil tip or a small screw driver. Status indication

adjustment (”Set”) is displayed.

➨ Push key-button again in order to adjust pressure reading to atmospheric pressure. Status indication

adjustment vanishes.

Instructions for repair with directions for repair and spare parts list is available on
request.
☞ The instructions are for trained service people.