Safety information – VACUUBRAND VAP 5 User Manual

Page 4

Advertising
background image

page 4 of 17

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 999039 / 16/12/2009

Safety information!

Remove all packing materials, remove the equipment from its packing-box, remove
the protective covers and inspect the equipment.
If the equipment is damaged, notify the supplier and the carrier in writing within three
days; state the item number of the equipment together with the order number and the
suppliers´ invoice number. Retain all packing materials for inspection.
Do not use the equipment if it is damaged.

☞ Read this manual before installing or operating the equipment.

Obey national safety regulations and safety requirements concerning the use of
vacuum and electrical equipment.
☞ Product must be connected only to a suitably fused and protected electrical supply

and a suitable earth point. Failure to connect the device to ground may result in
deadly electrical shock.

☞ Check that mains voltage and current conform with the product (see rating plate).

☞ Ensure that installation is in compliance with limitations from the degree of protec-

tion, see “Technical data“.

☞ The product is equipped with a short circuit proof transformer with an integrated

overload protection (no fuses).

Operate the equipment only in combination with VACUUBRAND genuine accessories.
Obey all other relevant safety requirements (e. g. accident prevention regulations for
working in laboratories and explosion protection guidelines) and provide protective
measures
.

Max. ambient temperature: 40°C
☞ Make sure ventilation is adequate if product is installed in an enclosure or if ambi-

ent temperature is elevated.

The devices are not suitable when working with dangerous or explosive gases or
explosive or flammable mixtures
. Ensure that the materials of the wetted parts are
compatible (see “Technical data“). If necessary adopt suitable measures.
☞ Temperature of the Pirani sensor is about 170°C.

☞ If residues can occur or if aggressive or condensable substances may enter the

product, install suitable protection (e. g. gas washing bottle).

☞ Position the product in the vacuum system so as to avoid flow of condensate

towards the gauge head.

DVR 5:
Max. permitted pressure:
1.5 bar (absolute).
☞ At a pressure greater than 1060 mbar the digital pressure display and the warning

triangle are flashing.

Attention: At a pressure greater than 1100 mbar the pressure measurement is incor-
rect (saturation limit of the pressure transducer).

VAP 5:
Max. permitted pressure:
1.5 bar (absolute).
Exclude sudden pressure changes when gauge head is operated.
☞ Pirani filament can be destroyed.
Ensure that the gauge head is electroconductively connected to an earthing wire when
the Pirani sensor is exposed to charged particles (ion-beam, plasma).
☞ Use only electroconductive centring/clamping rings. Do not connect the gauge head

via a nonconductive hose.

Advertising
This manual is related to the following products: