Technical data, See „technical data“, pg. 12) is not exceeded, And gas temperatures (see „technical data“, pg – VACUUBRAND VNC 2 User Manual

Page 12: 12), especially, Section „technical data“, pg. 12

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999161 / 27/03/2012

page 12 of 45

* The actual available range of the vacuum control can be limited due to the ultimate vacuum of the

pump, the developed amount of gas, etc.

** if pumping potentially explosive atmospheres: +10°C to +40°C

*** ohm resistive (inductive) load

Technical data

Controller

VNC 2 (E)

(not for

VACUUBRAND

NT VARIO pump or

pumping unit)

VNC 2 VARIO E

(only for connection

to VACUUBRAND

NT VARIO pump or

pumping unit)

Pressure transducer

capacitive, absolute pressure transducer made of

aluminium oxide ceramics, independent of gas type

Display

alphanumeric LCD display, illuminated

2 x 8 characters

Pressure units / scale (selectable)

mbar, Torr or hPa

Measuring range (absolute)

1100 mbar - 1 mbar (825 Torr - 1 Torr)

Maximum range of pressure control (absolute)*

1060 mbar – 1 mbar (795 Torr – 1 Torr)

Maximum permissible pressure at pressure

transducer (absolute)

1.5 bar (1125 Torr)

Measurement uncertainty (absolute) after careful

adjustment and at constant temperature

<

±1 mbar (0.75 Torr) ±1 digit

Temperature coefficient

<

± 0.07 mbar/K (0.05 Torr/K)

Ambient temperature range (operation)

10°C to +40°C

Ambient temperature range (storage)

-10°C to +70°C

Permitted relative atmospheric moisture during

operation (no condensation)

30% to 85%

Maximum permissible temperature of gaseous

media at pressure transducer

continuous operation: 40°C**

for short periods (< 5min) up to 80°C**

Max. permitted range of supply voltage

100 V~ (-10%) to 230

V~ (+10%) 50/60 Hz

24 V DC

±10%

(VACUU•BUS)

Fuse (IEC connection)

5x20 mm, T 8 A / 250 V

Power draw (no-load operation)

3 VA

2 VA

Power draw with VACUUBRAND VV-B 6C in-line

valve and coolant valve VKW-B

max. 10 VA (without

switch output)

-

Serial interface

RS 232 C

Supply line VACU•UBUS

voltage supply 110-230 V 50/60 Hz

voltage supply 100-110 V 50/60 Hz

24 V, max. 400 mA

24 V, max. 340 mA

-

designed for the simultaneous operation (parallel

connection) with the following original accesso-

ries:

one in-line valve VV-B 6C 24V=

one coolant valve VKW-B 24 V=

one venting valve VB M-B 24 V=

Breaking capacity of IEC socket***

ambient temperature 30°C

ambient temperature 40°C

7.2(4) A 250 V~ / 7.2(7.2) A 125 V~

6(4) A 250 V~ / 6(6) A 125 V~

Advertising