Working with the controller, See „working with the controller“, pg. 23, Pg. 23 – VACUUBRAND VNC 2 User Manual

Page 23

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999161 / 27/03/2012

page 23 of 45

Program menu in basic mode pControl

pControl
p Set

pControl

∆p

pControl
t Off

p Set
25mbar

∆p
Auto

t Off
off

1013mbar
Start

Working with the controller

Factory-set mode ”Vacuum control” (pControl)

Control without in-line valve:

• Control of a vacuum pump by direct switching of the pump. A preset time delay has an effect only on

an optional coolant valve.

Control with in-line valve:

• Control of a vacuum pump via an in-line valve. The pump continues to run according to the preset

time delay. This can considerably reduce possible condensation inside the pump. The coolant valve

is open during vacuum control and during time delay.

• For synchronous switching of the pump and the in-line valve:

Configure the time delay in the device menu: Delay: off.

Operation of a speed controlled NT VARIO vacuum pump (only with VNC 2 VARIO E):

• If after switching on the pump (without in-line valve) a NT VARIO pump is detected, the controller

changes automatically to the configuration ”NT VARIO” and stores this configuration. After a restart

the configuration ”NT VARIO“ is active directly.

Use keys s or t in the program menu to select an item.

Confirm selection with key ”Enter”.

Adjust parameters with keys s or t.

Confirm with key ”Enter” or return to original setting pressing ”Esc”.

pControl
Limit

Limit
1060mbar

Esc