Accu-Chek Aviva Combo User Manual

Page 236

Advertising
background image

234

Display Shows Possible Cause(s)

Possible Solution(s)

Connection­Lost

j

­­The­connection­was­lost­

between­the­meter­and­the­
pump­during­the­pairing­
process.­Therefore,­the­
attempt­to­pair­the­meter­and­
pump­was­unsuccessful.

j

­­Restart­the­pairing­process.­See­the­Advanced­

Owner’s­Booklet.

j

­­Pump­is­out­of­range­of­the­

meter.

j

­­Ensure­the­meter­and­pump­have­Bluetooth wireless­

technology turned­on.

j

­­Ensure­the­meter­and­pump­are­within­

communication­range.

j

­­If­the­low­battery­icon­is­displayed­on­the­meter,­the­

meter­cannot­communicate­with­the­pump.­Replace­
the­batteries.

Diary­Entry­Used­
For­Advice

j

­­The­diary­entry­selected­has­

been­used­for­bolus­advice­
and­modifications­are­not­
allowed.

j

­­Carefully­confirm­all­information­involving­bolus­

advice.­You­cannot­modify­bolus­advice­related­
entries­on­the­meter.­

Diary­Results­
Expired

j

­­­Bolus­advice­is­no­longer­

available­for­this­result.

j

­­Review­data­that­was­saved­in­My­Data.­Then,­

decide­to­modify­data,­add­data,­or­begin­a­new­
bolus­advice­session.

50813_00048009694_D.indb 234

7/3/12 3:13 PM