Display Shows Possible Cause(s)

Possible Solution(s)

Invalid­Target­
Range

j

­­The­lower­target­range­value­

is­above­the­upper­target­
range­value.

j

­­Reset­the­target­range­values.

j

­­Enter­the­correct­lower­target­range­value­and­

upper­target­range­value.­See­Chapter­5,­“Changing­
Meter­Settings.”

j

­­The­range­selected­conflicts­

with­your­hyper­and/or­hypo­
warning­limit­settings.

j

­­Reset­range­or­revise­warning­limit­settings­and­

re-enter­range.

j

­­Enter­the­correct­lower­target­range­value­and­

upper­target­range­value.­See­Chapter­5,­“Changing­
Meter­Settings.”

LO­bG­Warning

j

­­Your­blood­glucose­may­be­

lower­than­the­measuring­
range­of­the­system.

j

­­Treat­your­low­blood­glucose­as­recommended­by­

your­healthcare­professional.

j

­­If­you­are­experiencing­any­of­the­common­

symptoms­of­low­blood­glucose,­contact­your­
healthcare­professional­immediately.

Meter­and­Pump­
Not­Paired

j

­­Meter­and­pump­have­not­

been­paired.­You­cannot­use­
the­pump­functions­without­
pairing­the­meter­and­the­
pump.

j

­­You­must­pair­the­meter­and­pump­to­use­these­

features.

j

­­For­instructions­on­how­to­pair­the­meter­and­pump,­

see­the­Advanced­Owner’s­Booklet.

Meter­Battery­Low

j

­­Battery­power­is­low.

j

­­Install­new­batteries.­See­Chapter­7,­“Care­and­

Maintenance.”