Elenco Digital Bench Multimeter User Manual

Page 15

Advertising
background image

DVM645BI GB

14

Red

Circuit

Measuring Frequency and Adaptive Measurement

















Figure 2-5. Measuring Frequency and Adaptive Measurement (ADP)

c

Turn the selector to frequency and adaptive measurement.

d

Connect the leads as shown

e

is not available in frequency and adaptive measurement, all other buttons

can be used.

Each press the button

switches alternately between frequency and adaptive

measurement.

1. Insert the black and red test leads into the COM and V

Hz input terminals

respectively.

2. In the frequency test, it is not necessary to know the range, as the multimeter

will automatically range up or down to display the best resolution. Manual range is

not available.

3. Ensure that the amplitude level of the signal to be measured is not greater

than the input voltage limit (250V DC/AC rms). The signal amplitude must also be

greater than the sensitivity level.

4. Attach the probe tips to the points across which the frequency is to be

measured, and read the result directly from the display.

5. The frequency range is always set to auto range.

The Adaptive (ADP) Measurement is as follows:

1. Connect the ADP signal terminal (-) and (+) to the COM and V

Hz input

terminals respectively.

2. The ADP voltage is supplied to the multimeter directly. The result is displayed

on a scale of 10 units per 1mV.

3. The ADP measurement is always fixed range. The full range is 400mV DC.