56

APPLICATION

PHASE DIFFERENCE MEASUREMENTS

Method No.1

(Refer to Fig.15)

This procedure is useful in measuring the phase difference of signals of the same
frequency.

1. Apply the two signals to the CH 1 and CH 2 input jacks, selecting the dual-trace

display mode (either ALT or CHOP).

2. Set the trigger SOURCE switch to the signal witch is leading in phases (or to LINE

if device is line voltage operated) and use VOLTS/DIV controls to adjust the two
waveforms so they are equal in amplitude.

3. Use the Y POSITION control to position the waveforms in the vertical center of the

screen. Use the TIME/DIV and VAR SWEEP controls to adjust the display so one
cycle of the reference signal occupies 8 divisions horizontally (see Fig.

). The

TRIG LEVEL and X POSITION controls are useful in achieving this display. The
display should be as shown in Fig.

, where one division now represents 45 in

phase.

4. Measure the horizontal distance between corresponding points on the two

waveforms. Multiply the distance (in division) time 45 per division to obtain the
phase difference.

The measurement is summarized by the following equation:

Phase difference = Hor div x 45 /div

For the example shown in Fig.15, the horizontal distance is 1.7 divisions. Thus,
the phase difference is calculated as follows:

Phase difference = 1.7 x 45 /div = 76.5

15

15

°

°

°

°

°

Fig.15
Phase Difference Measurement

100

90

10

0

1 cycle

Phase difference

Comparison
signal

Reference Signal