4.0 Operation

When the

CP-10

is installed as described in Section 3, and in clean air, the

POWER

green LED is on, the display is lit and the

information on the display is measurement of the target detected by the

CP-10.

The red alarm and fault LEDs are not lit.

4.1 Start Up CP-10

When the

CP-10

is first powered up, it goes through a series of momentary screens, which identify the instrument model number,

serial number and software revision. After all of the momentary screens have been displayed, the instrument arrives at the Main
Gas Display showing the gas concentration and unit of measurement in ppm, % or %LEL.

Depending on transmitter configuration and calibration condition, the furthest right character in the display may flash a letter
indicating the instrument status. See the Section 4.1.1 below

4.1.1 Typical Start Up

When power is supplied to the

CP-10

, the instrument will display the following sequence of information:

Typical start up sequence of information displayed.

Example of Typical Start Up Display

Function

The instrument: Model

CP-10

Example for reference only

The instrument: Serial Number

Example for reference only

The instrument: Software Revision

IF the right most character is a flashing

W

The instrument is in Warm-up mode

This should last about 1 minute

The Signal Output is held at 4mA during warm-up

The instrument: Normal Display Mode
Measurement of the target gas

PPM Parts Per Million
%

Percent By Volume

LEL Lower Explosive Limit
F

Fault, Sensor/Transmitter is not connected or

wiring is not correct

N

OTE

:

Software revision may cause variations of display output.

CP-10

130-1256

S/W X.X

0ppW

0ppm

0LEL

20.9%