420setup, Ma trim scale4ma, Scal20ma – ENMET CP-10 User Manual

Page 14: Cal ok, Cp-10 maintenance menu flow chart – oxygen

Advertising
background image

CP-10

ENMET Corporation

12

Normal Gas Display

20.9%

Menu Press and H

OLD

the M

ENU

button for 2 – 4 seconds to enter the Maintenance Menus

Exit

Select

Press the Select button to return to the Normal Gas Display.

See Section 5.2.1

Menu

Press the Menu button to cycle through Maintenance Menus

mA Span

Menu

Select

30.0

To change mA Span set point:
Press Menu button until mA Span is displayed
Press Select button to display the set point

The Menu button changes digit indicated by underscore cursor
The Select button locks underscored digit and moves to next digit

See Section 5.2.4

To return to Normal Gas Display:
Press M

ENU

button until EXIT is displayed

Then press S

ELECT

button

To change Alarm set points:
Press Menu switch until Alarm to be changed is displayed
Press Select switch to display the set point

The M

ENU

switch: changes digit indicated by underscore

cursor
The S

ELECT

switch: locks in the underscored digit and

moves to next digit

If change is not within range display returns to first digit
If change is within range display moves to Set Time Delay

Use M

ENU

and S

ELECT

switches as above to change time

delay. Between 0 and 5 seconds is allowed

If change is within range display moves to next menu

See Section 5.2.3

Λ

- Indicates increasing alarm

V - Indicates decreasing alarm

SetTDsec

0

Alarm3

S

ELECT

Λ 23.5

M

ENU

SetTDsec

0

Alarm1

M

ENU

S

ELECT

V

17.0

SetTDsec

0

Alarm2

M

ENU

S

ELECT

V

19.5

420Setup

Menu

Select

Select

Select

mA Trim

Scale4mA

0.0

Scal20mA

30.0

20.9

Cal OK

Select

Menu

Press Select button to set low end of scale:

The Select button locks underscored digit and moves to next digit
The Menu button changes digit indicated by underscore cursor
When last digit is entered CP-10 will advance to full scale set up

Select

Menu

Menu

Select

Press Select button to advance to display trim
To adjust display of CP-10 to match S/T display

The Select button locks underscored digit and moves to next digit
The Menu button changes digit indicated by underscore cursor

When the last digit has been locked in place, Cal OK will appear briefly than
the CP-10 will advance to the next menu, Alarm 1

See Section 5.2.2

Press Select button to advance to full scale
To set full scale :

The Select button locks underscored digit and moves to next digit
The Menu button changes digit indicated by underscore cursor
When last digit is entered CP-10 will advance to display trim

N

OTE

:

Software revision may cause variations of display output.

F

IGURE

7:

CP-10 Maintenance Menu Flow Chart – Oxygen