0 features, Figure 1: front view – ENMET GS-24-DF Gas Sampler User Manual

Page 6

Advertising
background image

2

2.0 Features

See Figure 1 and 2

Feature

Description

Power Terminal

This terminal provides connections for: the DC power (24 V

DC

) supplied from external

power source, Sensor/Transmitters and 4-20mA

Display

LCD, Allows the user to verify and monitor the flow rate of the air sample. Standard flow
rate, approximately 05 lpm(liter per minute).

Indicator LED

Green/Red: When green indicates flow rate is sufficient.

When red indicates flow rate is inadequate.

Yellow: When on solid indicates GS-24-DF is in menu to change settings

When flashing indicates GS-24-DF is in menu to calibrate(factory use only)

Sampling Pump

This electromagnetic diaphragm pump draws the air sample from the test area to the sensor
chamber.

Pushbutton Switch

SW1: Menu Switch, this switch is for viewing/changing display settings.
SW2: Select Switch, this switch is for temporary deactivating audio alarm and changing

display settings.

Potentiometer

Flow Adjustment Potentiometer (POT)

Inlet Port

This port draws the air sample from the test area. For external piping, use ¼

″ O.D. tubing.

See section 4 for type of tubing to be used.

Outlet Port

This port expels the air sample after it passes the sensor. For external piping, use ¼

″ O.D.

tubing

Gas Sensing Chamber

This chamber directs the air sample to the gas sensor. Varies, depending on the type of
sensor/transmitter.

Figure 1: Front View

Flow Rate Indicator Red when in Alarm

Menu Indicator Yellow when in Menu Mode

Power Terminal

Pushbutton switch M

ENU

(SW1)

Pushbutton switch S

ELECT

(SW2)

Outlet Port

Inlet Port

Gas Sensing
Chamber

Potentiometer for
Flow Adjustment