6.2 Pump Maintenance for Pump 04018-118

For the maintenance of ENMET part number 04018-118.

Top View

Side View

Figure 10: 04018-118 Pump Features

6.2.1 Pump Diaphragm Replacement

If the diaphragm becomes damaged or torn, it can be replaced.

Replacement diaphragm ENMET part number 04018-117.

Figure 11: Diaphragm Details

Remove 2 star head screws and pump arm. See figures 10 and 11.

Remove damaged diaphragm from pump arm and replace with new diaphragm.

Reattach pump arm and verify diaphragm is properly sealed to check valve housing.

Pump Arm

Check Valve

Housing

Diaphragm

Mounting

Screws

3 places

Star Head
Screw

2 places