SCPI Command Reference

126

IQS-3150

Product-Specific Commands—Description

DEFault allows the instrument to select a value

for the <Offset> parameter.

The <Offset> parameter corresponds to a valid

power offset in dB. The offset value will be added

to the absolute power. The same offset value will

be used for all wavelengths. You can use the

OUTPut[1..n]:OFFSet? MAX and

OUTPut[1..n]:OFFSet? MIN queries to determine

a valid range for the power offset.

Example(s)

INP:WAV 1310 NM

CONT:MODE POW

OUTP:ALC:STAT OFF

OUTP:APM ABS

OUTP:OFFS 0.000 DB

OUTP:POW -5.500 DBM

OUTP:POW? Returns: -5.500000E+000

OUTP:RPOW? Returns: -5.500000E+000

OUTP:OFFS -1.500 DB

OUTP:POW? Returns: -5.500000E+000

OUTP:RPOW? Returns: -7.000000E+000

OUTP:OFFS MAX

See Also

:INPut[1..n]:OFFSet

:OUTPut[1..n]:OFFSet?

:OUTPut[1..n]:RPOWer

:OUTPut[1..n]:OFFSet