EXFO IQS-600 Integrated Qualification System User Manual

Page 262

Advertising
background image

SCPI-Based Errors

254

IQS-600

–233

“Invalid version”

Indicates that a legal program data element was

parsed but could not be executed because the

version of the data is incorrect to the device. This

particular error should be used when file or block

data formats are recognized by the instrument but

cannot be executed for reasons of version

incompatibility. For example, file or instrument

version that are not supported.

–240

“Hardware error”

Indicates that a legal program command or query

could not be executed because of a hardware

problem in the device. The definition of what

constitutes a hardware problem is completely

device-specific. This error message should be used

when the device cannot detect the more specific

errors described for errors –241 through –249.

–241

“Hardware missing”

Indicates that a legal program command or query

could not be executed because of missing device

hardware; for example, an option was not

installed. The definition of what constitutes missing

hardware is completely device-specific.

–250

“Mass storage error”

Indicates that a mass storage error occurred. This

error message should be used when the device
cannot detect the more specific errors described

for errors –251 through –259.

–251

“Missing mass storage” Indicates that a legal program command or query

could not be executed because of missing mass

storage; for example, an option that was not

installed. The definition of what constitutes missing

mass storage is device-specific.

–252

“Missing media”

Indicates that a legal program command or query

could not be executed because of a missing

media; for example, no disk. The definition of what

constitutes missing media is device-specific.

Error

Number

Description

Probable Cause

Advertising