Hanna Instruments HI 98160 User Manual

Page 11

Advertising
background image

11

pppppH CALIBRATION

H CALIBRATION

H CALIBRATION

H CALIBRATION

H CALIBRATION

Hanna recommends frequent calibrations, especially if high accuracy is

required.

The pH range should be recalibrated:

• Whenever the pH electrode is replaced.

• At least once a week, but daily is recommended.

• After testing aggressive chemicals.

• After electroded is cleaned.

• When calibration alarm time out is expired - “Cal Due” tag blinks (if

feature is enabled).

• If “Outside Cal Range” message blinks during pH measurement

(the measurement range is not covered by current calibration).

PROCEDURE

The instrument offers a choice of 7 buffers that are automatically recognized

(1.68, 4.01, 6.86, 7.01, 9.18, 10.01 and 12.45 pH) and also allows

the user to enter two more pH values for calibration, “Custom 1” and

“Custom 2”. The set custom buffers are the buffer values at 25 ºC.

HI 98160 offers also 3.00 pH buffer choice for wine measurement.

When a custom buffer is selected during calibration, the “Custom” tag

is displayed on the LCD and its value can be changed in a ±1.0 pH

window, around the set value, in accordance with current temperature.

For accurate pH measurements, it is recommended to perform a five-point

(HI 98150 & HI 98160) calibration respectevely a three-point (HI 98140)

calibration. However, at least a two-point calibration is suggested.

The instrument will automatically skip the buffers used during calibration

and the buffers which are in a ±0.2 pH window, around one of the

calibrated buffers.

• Pour small quantities of selected buffer solutions into clean beakers.

For accurate calibration use two beakers for each buffer solution, the

first one for rinsing the electrode and the second one for calibration.

• Remove the protective cap and rinse the electrode with some of the

buffer solution to be used for the first calibration point.

FIVE-POINT CALIBRATION (HI 98150 & HI 98160)

• Submerse the pH electrode and the temperature

probe approximately 3 cm (1¼”) into a buffer

solution of your choice (pH 1.68, 4.01, 6.86,

7.01, 9.18, 10.01, 12.45 or a custom

buffer) and stir gently. The temperature

probe should be close to the pH electrode.

• Press CAL. The instrument will display the

measured pH on the primary LCD and “7.01”

buffer on the secondary LCD, together with

“CAL” and “Cal Point 1” tags.

3cm

(1¼”)

Advertising