The LCD will show the pH measurement

together with the temperature of the

sample. In order to take more accurate

pH measurements, make sure that the

electrode and instrument is calibrated (see

page 11 for details).

Hanna recommends to keep the pH electrode wet and to rinse throughly

with the sample before measurements.

The pH reading is directly affected by temperature. In order for the

instrument to measure the pH accurately, temperature must be taken

into consideration. If the sample temperature is different from the

ambient temperature, allow a few minutes for the pH electrode to reach

thermal equilibrium.

The HI 1618D pH sensor contains an integral temperature sensor so the

pH reading is automatically temperature compensated (ATC) on the

HI 98140 and HI 98150. If the temperature is out of range (due to

temperature sensor failure) these instruments will enter MTC mode and

permit temperature values to be set manually using the ARROW keys.

To use the instrument's Automatic Temperature Compensation feature for

HI 98160, connect and submerse the temperature probe into the sample

as close to the electrode as possible and wait for a few seconds.
If manual temperature compensation (MTC) is desired, the

temperature probe must be disconnected from the instrument

(HI 98160 only).
The display will show the default temperature

of 25 ºC, the last measured temperature

reading, or the last set temperature, with

the “ºC” (or “ºF”) tag blinking.
The “MTC” tag and

arrow symbols

light up on the LCD to indicate that the

instrument is in MTC mode and the arrow

keys can be used to enter the desired temperature value.
Note: When in MTC the user can press and hold the ARROW keys to

set the measurement temperature. The instrument will start

incrementing/decrementing the temperature value. The instrument

keeps measuring using the MTC value for temperature compensated

pH and the display is updated.

ORP MEASUREMENTS (HI 98150 & HI 98160 only)
To perform ORP measurements, connect an optional

ORP electrode (see Accessories section) to the instrument

and turn it ON. If necessary, enter the mV mode by

pressing Range until the display changes to mV.