Hanna Instruments HI 83200-01 User Manual

Page 56

Advertising
background image

110

111

Potassium MR

POTASSIUM MEDIUM RANGE

SPECIFICATIONS

Range

10 to 100 mg/L

Resolution

2.5 mg/L

Accuracy

±15 mg/L ±7% of reading

Typical EMC

±2.5 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 610 nm

Method

Adaptation of the Turbidimetric Tetraphenylborate method. The reaction between

Potassium and reagents causes turbidity in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93750A-0

Potassium Reagent

6 drops

HI 93750B-0

Powder Reagent

1 packet

REAGENT SETS

HI 93750-01 Reagents for 100 tests
HI 93750-03 Reagents for 300 tests
For other accessories see page 128.

MEASUREMENT PROCEDURE

Note: for sample preparation follow the NUTRIENTS SOLUTION (MR) procedure at page 18.

• Select the

Potassium MR method using the procedure

described in the

Method Selection section (see page 12).

• Fill the cuvette with 10 mL of sample, up to the mark.

• Add six drops of HI 93750A-0, replace the cap and swirl

the solution.

• Place the cuvette into the holder and close the lid.

10 mL

x 6

Potassium HR

POTASSIUM ULTRA HIGH RANGE

For samples containing more than 200 ppm of Potassium: follow the sample preparation procedure
described at page 19 for NUTRIENTS SOLUTION (HR). Then add to the graduated cylinder 20 mL of the

prepared sample (for HR) and fill the cylinder with demineralized water from the Demineralizer Bottle up

to the 100 mL mark.
Follow the MEASUREMENT PROCEDURE at page 108.
Read the result in mg/L of Potassium on the display and multiply the reading by 5 to obtain the actual

concentration of Potassium.

INTERFERENCES

Interferences may be caused by:
Ammonium above 10 ppm
Calcium above 10000 ppm as CaCO

3

Chloride above 12000 ppm
Magnesium above 8000 ppm as CaCO

3

Sodium above 8000 ppm