Hanna Instruments HI 83200-01 User Manual

Page 59

Advertising
background image

116

117

Silica

• Place the cuvette into the holder and close the lid.

• Press ZERO key. The display will show “-0.0-” when the

meter is zeroed and ready for measurement.

• Remove the cuvette and add the content of one packet

of HI 93705C-0 Amino acid reagent and shake until it is

completely dissolved.

• Reinsert the cuvette into the instrument.

• Press TIMER and the display will show the countdown

prior to the measurement. Alternatively, wait for exactly 3
minutes and press READ. When the timer ends the meter

will perform the reading. The instrument displays

concentration in mg/L of silica (SiO

2

).

• Press the or to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of Silicon (Si).

• Press the or to go back to the measurement screen.

INTERFERENCES

Interference may be caused by:
Phosphate above 60 mg/L
Phosphate above 75 mg/L
Sulfide and high concentration of iron
Eliminate color and turbidity interferences by zeroing the meter with the original water sample.

Silica

Advertising