Hanna Instruments HI 4211 User Manual

Page 23

Advertising
background image

2 3

• Press

to exit custom buffer edit menu. If the Saving Confirmation is enabled, press

to accept

the modified option,

to escape without saving or

to return to the editing mode. Otherwise,

the modified option is saved automatically.

• Use

key to select the next custom buffer to be set or press

to return to Calibration options.

Edit Buffer Group
Accessing this option the user can edit the desired group of five pH buffers for automatic buffer recognition
(Automatic Buffer Entry Type). If the Buffer Group already contains five pH buffers, at least one pH buffer has to
be removed in order to add another buffer.

To edit/set the Buffer Group:
• Press

while in

pH Measure mode.

• Press

.

• Use

or

to select the Calibration option.

• Press

and use

or

to highlight the Edit

Buffer Group option.

• Press

and use

and

to choose the pH

buffer to be included in the buffer group.

• Press

or

to add/remove the selected pH buffer

to/from the buffer group.

• Press

to return to Calibration options and to save the changes.

Calibration Reminder
In order to have accurate readings, the instrument must be calibrated frecquently. Three options are available for
the calibration reminder: Daily, Periodic or Disabled.

To set the Calibration Reminder:
• Press

while in

pH Measure mode.

• Press

.

• Use

or

to select the Calibration option.

• Press

and use

or

to highlight the

Calibration Reminder option.

• Press

and use

or

to highlight the

desired option.

• Press

to confirm your selection or press

to

cancel operation.