Hanna Instruments HI 4211 User Manual

Page 37

Advertising
background image

3 7

• To select the appropriate Electric Charge/Slope use

or

and then press

. If the Ion

electric charge is None, its slope can be manually set by
pressing

. A pop-up menu will be displayed on

the LCD, in which the slope value can be set using
or

. Press

to save the modified value or

press

to return to the Ion Constants menu.

Note: If an ISE calibration was performed and a different Ion

Selective Electrode is selected (standard or custom), a
warning message appears on the LCD informing the user to
perform a new calibration or to select the previous ISE in
order to perform accurate measurements.

Concentration Unit
Accessing this option, the user can select the desired concentration unit for the measured ion or chemical compound.
The available concentration units are: ppt (g/L), ppm (mg/L), ppb (μg/L), mg/mL, M (mol/L), mmol/L and %w/v.
To set the Concentration Unit:
• Press

while in

ISE Measure mode.

• Press

.

• Use

or

to select the Concentration Unit option.

• Press

and use

or

to highlight the

desired option.

• Press

to confirm your selection or press

to

cancel operation.

Sample ID – See pH Setup section, page 20.

Stability Criteria – See pH Setup section, page 20.

Log – See pH Setup section, page 20.
Note: The Logging Data Configuration option includes also the Ion Constants parameter. If you want it to appear

in the log reports, it must be enabled.

Alarm – See pH Setup section, page 20.
Note: The Alarm Limits (Low and High) are set in the selected concentration unit of the measured ion or chemical

compound and are displayed in a scientific format (value and value exponent).