Hanna Instruments HI 9813-5 User Manual

Page 11

Advertising
background image

11

Accessories needed:

• Use

HI 70031 (1413

µ

S/cm = 1.413 mS/cm) EC calibration

solution or

HI 70442 (1500 ppm) TDS calibration solution.

Note: The conversion between EC and TDS is made by a built-in

circuit, hence it is requested to calibrate the meter only in EC
or TDS range. The other range is thus automatically calibrated.

PROCEDURE
• Pour approximately 4 cm (1½") of a conductivity calibration

solution (e.g.

HI 7031) into a beaker. If possible, use plastic

beaker to minimize any EMC interference.

• Immerse the probe in the solution.
• Wait for a couple of minutes for thermal equilibrium to be

reached.

• Tap the probe on the bottom, then shake it lightly while rotating

to make sure no air bubbles remain trapped inside the probe.

• Press the

mS/cm (or ppm) key.

• Turn the EC/TDS calibration knob until the

display shows the EC or TDS reading at
25°C.