CLEANING PROCEDURE
For better accuracy in measurements and to ensure a good performance
of the probe, a frequent cleaning is recommended.
For this purpose, soak it in Hanna

HI 700661 Cleaning Solution for

5 minutes.
Note: For particular dirty (as for example protein, oil or grease) see

the “Accessories” section for Hanna specific solutions.

HI 9813-6 only:

CAL-CHECK

TM

FEATURE / CLEAN & CAL PROCEDURE

The unique Cal-Check

TM

feature allows the user to check the meter

calibration status at any time. Simply follow the below procedure for
a quick and easy operation.
• Rinse the probe with water.
• Immerse the probe in the

HI 50021 check solution.

Note: If the reading is around pH 4.0, the probe is broken and has

to be replaced.

• Press the

Check key.

• If the meter is calibrated “

Probe is

OK” message is displayed.

• If cleaning is needed “

Clean Probe

and Calibrate” message is displayed:
1. immerse the probe in the

HI 700661 cleaning solution (or
in one of the available specific
solutions - see “Accessories” section)
for 5 minutes, then rinse it with tap water.

2. immerse again in the

HI 50021 check solution and press the

CHECK key.

3. if the message “

Probe is OK” is displayed, the meter does not

need calibration.

4. if the message “

Clean Probe and Calibrate” is displayed

again, calibration is needed (see “pH Calibration” section).

Notes: • If it is not possible to calibrate, the probe has to be

replaced with a new one.

• For field applications, it is always recommended to keep a

spare probe handy. When anomalies are not resolved with
simple maintenance, change the probe and recalibrate
the meter.