Hanna Instruments HI 98710 User Manual

Page 7

Advertising
background image

13

12

TO RESET PRINTING SAMPLE NUMBER

(HI98710 and HI98740 only)

This feature resets the print on demand sample number to 001.
• Select the code 04
• A scrolling message will be displayed.
• Press the ALT and CFM keys to confirm reset or ALT and FNC to

escape.

GLP DATA

Cod. 10 - viewing GLP data
• Select the code 10
• A message will scroll twice across LCD.
• The LCD will then display the instrument identification (ID) code.
• Press to scan remaining data, in the following order:

channel 1 last calibration date (DD.MM)
channel 1 last calibration year
channel 2 last calibration date (DD.MM)
channel 2 last calibration year
channel 3 last calibration date (DD.MM)
channel 3 last calibration year
channel 4 last calibration date (DD.MM)
channel 4 last calibration year

Note: Data can be viewed in reverse order pressing the key.
• Press ALT and FNC to return to function selection mode.
Code 20 - setting the identification (ID) code
When using several identical meters it may be useful to uniquely
identify them by assigning an ID code to each meter.
• Select code 20. A message will scroll across LCD.
• Enter a 4-digit value using the arrow keys.
• Press ALT and CFM to confirm the value.

TESTING BATTERY LEVEL

• Select code 41. The message "Battery test" will scroll across LCD.
• If the meter is connected to an external power adapter, the LCD

will display "LINE" otherwise it will display bAtt on the upper
display, and the remaining percentage of battery charge (100%
means fully charged battery and 0% cor-
responds to the minimum battery voltage
that allows the meter to operate).

Viewing logged data
• Press CH/TIME (HI98840 only) or TEMP/TIME (HI98810 only) to

view data of the selected sample. Data will be displayed in the
following order:

channel 1 temperature reading
channel 2 temperature reading (HI 98840 only)
channel 3 temperature reading (HI 98840 only)
channel 4 temperature reading (HI 98840 only)
date
time

• If CH/TIME or TEMP/TIME is pressed when the time is displayed,

the LCD will pass to the next Sample number.

• It is then possible to scroll the data of the next sample by pressing

CH/TIME or TEMP/TIME or select a different sample by the arrow
keys.

• to return to setup mode, press ALT and FNC when the meter

displays the sample number.

Cod. 02 - Lot summary printout
• Select the code 02.
• The message "Lot table Printing" will scroll twice across LCD.
• The meter will then print a complete set of information based on

the data stored in memory:

a - lot number
b - logging interval
c - starting date
d - number of samples
e - starting time
f

- logged channels numbers.

Note For lot 00 the number of samples only will be printed.

DELETE LOGGED DATA (HI98810 and HI98840 only)

• Select code 05 to delete the Log on Demand data or code 06 to

delete the Timed Log Data.

• A scrolling message will be displayed.
• Press the ALT and CFM keys to confirm deletion.
• It is also possible to escape without data deletion pressing the ALT

and FNC keys.