OPERATIONAL GUIDE

Each meter is supplied complete with a 9V battery. Remove the battery cover
from the back of the thermometer, insert the battery and reattach the cover.
Connect the temperature probe to the probe socket on the top of the meter.

To switch the instrument on, turn the rotary switch to
°C or °F position (depending on model).

If a temperature probe is plugged in, the unit will display the measured
temperature.

After use, switch the thermometer off by turning the
rotary switch to OFF.

CALIBRATION

All HANNA instruments

®

thermometers have been accurately pre-calibrated at

the factory. It is generally recommended to have all thermometers recalibrated
at least once a year. For an accurate annual recalibration, contact the nearest
HANNA Customer Service Center.

For a quick and easy check of the meter accuracy,
HANNA instruments

®

offers a series of CALIBRATION

TEST KEYS to be connected to the probe socket on
the thermometer.

If the reading differs more than ±0.4°C (±0.8°F) from the test key value, the
unit is due for recalibration.
Choose the test key that better fits your application needs:
HI 765-18C

Calibration test key at

–18.0°C

HI 765000C

Calibration test key at

0.0°C

HI 765070C

Calibration test key at

70.0°C

HI 765-004F Calibration test key at

– 0.4°F

HI 765032F

Calibration test key at

32.0°F

HI 765158F

Calibration test key at

158.0°F

OFF

°C

°C

°F

OFF

°C

°C

°F

FUNCTIONAL DESCRIPTION & SPECIFICATIONS

OF HI 8753

1) Probe connector
2) Liquid Crystal Display
3) OFF mode
4) Measurement mode, °C scale
5) Measurement mode, °F scale
6) Rotary switch

Range

–40.0 to 150.0°C / -58 to 338°F

Resolution

0.1°C / 1°F

Accuracy

±0.5% F.S. for one year, excluding probe error

Typical EMC Deviation

±3°C / ±6°F

with HI 765BL probe

Probe

HI 765BL stainless steel, 1 m (3.3') cable (included)

Battery Type / Life

1 x 9V (IEC 6LR61) / approx. 300 hours of continuous use

Environment

0 to 50°C (32 to 122°F);

RH max 95% non-condensing

Dimensions

180 x 83 x 40 mm (7.1x3.3x1.6")

Weight

265 g (9.3 oz.)