19

A 1/2 BSP sample input fitting allows direct connection to the
optional input filter.

Sample line pressure should be between 0.07 and 4 bar (1
and 57.2 psi) with an ideal pressure of 0.7 bar (10 psi).

It is recommended to assure that the water inlet come at 1 m
above the instrument sample input pitting. For maintenance
issues is also recommended to install valves on both sam-
pling pipes.

Drain Line Installation

The drain hose fitting is a 20 mm (3/4”) hose barb on the
bottom of the instrument enclosure. An air gap between the end
of the drain hose and the drain is recommended to prevent any
back flow into the instrument in the event of drain blockage.

ORP probe

PH probe

PCA 301

REAGENTS

HI 70460

TOTAL CHLORINE INDICATOR SOLUTION

R: - -
S: - -

ETICPCA30A

Contents:
p-Toluenesulfonic Acid (6192-52-5)
Water (7732-18-5)

FOR LABORATORY AND INDUSTRIAL USE ONLY
DO NOT FREEZE
STORE AT ROOM TEMPERATURE
KEEP CONTAINER TIGHTLY CLOSED

PCA 301

REAGENTS

HI 70461

TOTAL CHLORINE BUFFER SOLUTION

Xn: Harmful

R: 22
S: 46

ETICPCA30B

FOR LABORATORY AND INDUSTRIAL USE ONLY
DO NOT FREEZE
STORE AT ROOM TEMPERATURE
KEEP CONTAINER TIGHTLY CLOSED
HARMFUL IF SWALLOWED.
IF SWALLOWED SEEK MEDICAL ADVICE IMMEDIATELY

AND SHOW THIS CONTAINER OR LABEL.

Contents:
Lithium Bimaleate (unknown)
Lithium Maleate (unknown)
Water (7732-18-5)

TO DRAIN

FILTER

WATER INLET

WATER OUTLET

MIN.
1 m