Hanna Instruments PCA 310 User Manual

Page 73

Advertising
background image

73

“High ORP

High ORP

High ORP

High ORP

High ORP” on LCD and “H ORP

H ORP

H ORP

H ORP

H ORP” on SMS

The ORP is over the Alarm High setpoint.
Change the setpoint, verify the ORP probe.

“Low ORP

Low ORP

Low ORP

Low ORP

Low ORP” on LCD and “L ORP

L ORP

L ORP

L ORP

L ORP” on SMS

The ORP is under the Alarm Low setpoint.
Change the setpoint, verify the ORP probe.

“High T

High T

High T

High T

High Temperature

emperature

emperature

emperature

emperature” on LCD and “H T

H T

H T

H T

H Temp

emp

emp

emp

emp” on SMS

The Temperature is over the Alarm High setpoint.
Change the setpoint, verify the pH probe.

“LLLLLow T

ow T

ow T

ow T

ow Temperature

emperature

emperature

emperature

emperature” on LCD and “L T

L T

L T

L T

L Temp

emp

emp

emp

emp” on SMS

The Temperature is under the Alarm Low setpoint.
Change the setpoint, verify the pH probe.

WARNING MESSAGES

“Dirty Cell

Dirty Cell

Dirty Cell

Dirty Cell

Dirty Cell” on LCD and “Drt Cell

Drt Cell

Drt Cell

Drt Cell

Drt Cell” on SMS

The light level is too low. The light source current is near the
highest value. The cell is dirty or the water is too turbid.
Check the water quality; check the water flow; Clean the cell.

“Reagent Low Level

Reagent Low Level

Reagent Low Level

Reagent Low Level

Reagent Low Level” on LCD and “L Reag

L Reag

L Reag

L Reag

L Reag” on SMS

The reagent reach the 20% of the initial quantity. The re-
agent level is based on a “blind” counting of the com-
pleted measurements. A difference between reagent counter
and real reagent level could appear.
Change the reagent or reset the reagent counter.

“Cl Calibration Old

Cl Calibration Old

Cl Calibration Old

Cl Calibration Old

Cl Calibration Old” on LCD and “Cl Cal

Cl Cal

Cl Cal

Cl Cal

Cl Cal” on SMS

Cell calibration is older than 1 months.
Calibrate the cell.

“R

R

R

R

Reagent Expired

eagent Expired

eagent Expired

eagent Expired

eagent Expired” on LCD and “R

R

R

R

Reag Exp

eag Exp

eag Exp

eag Exp

eag Exp” on SMS

Reagent is older than 3 month. The reagent expiration
time is started when reagent counter is reset or “prime
reagent pump command is executed.
Change the reagent and reset the reagent counter.

“GSM Wrong PIN Code

GSM Wrong PIN Code

GSM Wrong PIN Code

GSM Wrong PIN Code

GSM Wrong PIN Code”: appears on LCD

Wrong PIN code was entered.
Set the correct pin code and restart the analyzer.

“GSM Init failed

GSM Init failed

GSM Init failed

GSM Init failed

GSM Init failed”: appears on LCD

The GSM modem was not successfully initialized.
Wait for auto retry. Verify other messages related to GSM.

“GSM Card Expired

GSM Card Expired

GSM Card Expired

GSM Card Expired

GSM Card Expired”: appears on LCD

The SIM expiration date is reached.

Advertising