Hanna Instruments HI 8614 User Manual

Page 3

Advertising
background image

5

4

SPECIFICATIONS OF HI 8614 & HI 8614L

HI 8614

pH TRANSMITTER

HI 8614

L

pH INDICA

TOR

& TRANSMITTER

pH

The transmitters use a universal BNC socket
for quick and secure connection to any elec-
trode with a BNC connector.

For HI 8614 and HI 8614L the temperature
compensation is performed by the transmitter's
ATC circuitry if the temperature probe
(HI 76608, optional) is attached. If ATC is not
required, it is possible to substitute the tem-
perature probe with a fixed resistor.

The input is isolated from the current loop to
eliminate problems related to ground loop,
low insulation cables, multiple electrode con-
nections, and a common mode voltage of up
to 100 V ensures true differential readings.

The terminal board of the transmitter provides
for connection of power supply, pH or ORP
electrodes and temperature probe (for HI 8614
and HI 8614L models only).

The unit is enclosed in a protective casing
conforming to IP 65 standards.

Range

4 to 20 mA / 0.00 to 14.00 pH

Resolution

0.01 mA / 0.01 pH

Accuracy (@20°C/68°F)

±0.02 mA / ±0.02 pH

Typical EMC Deviation

±0.25 mA / ±0.2 pH

Calibration

Offset: ±2.2 mA / ±2 pH

Slope: ±0.5 mA / 86 to 116%

Temperature Compensation

Fixed or automatic from 0 to 100°C

(32 to 212°F) with HI76608 probe

Input Impedance

10

12

Ohm

Output

4-20 mA (isolated)

Installation Category

I I

Protection

IP 65

Environment

0 to 50°C (32 to 122°F);

RH max 95% non-condensing

Power Supply

HI8614: 18 to 30 Vdc

HI8614L: 20 to 36 Vdc

Load

Max 500 Ohm

Dimensions

165 x 110 x 90 mm

(6.5 x 4.3 x 3.5'')

Weight

1 kg (2.2 lb.)