FOR HI 8615L ONLY

The HI 8615L is factory calibrated, and the
displayed values are referenced to the 4-20
mA loop current (e.g. LCD displays -1000
mV when loop current is 4.00 mA and dis-
plays +1000 mV when current is 20.00 mA).
Under normal application, adjustment on this
module may not be necessary. If routine
check is required, the following procedures
shall be performed.

Follow the above procedure for HI 8615
(see page 12).

Simulate a 12.00 mA loop current for the
transmitter (i.e. 0 mV at Electrode input)
and check display reading.

Simulate a 20.00 mA loop current for the
transmitter (e.g. +1000 mV at Electrode
input) and check display reading.

Set the simulator to 350 mV and adjust
the slope trimmer to read 14.8 mA on the
ammeter or 350 mV on the HI 8615L
display (HI 8615L only).

Connect the ORP electrode to the module
and immerse the tip of the electrode into
the beaker of HI 7020 ORP solution and
check that the ammeter reading lies be-
tween 13.6 and 14.2 mA or the instrument
reading is between 200 and 275 mV at
25°C (HI 8615L only).

Only if the reading lies out-
side this range, adjust the
slope adjustment trimmer on
the transmitter to reflect a
reading within this range.