Delphi DL-00183 User Manual

Page 14

Advertising
background image

2

DL-00096 Rev. 7

Charging the Battery

Prior to using the Portable Oxygen Concentrator for the first time, install the
battery as shown in Figure 1. The battery will latch when fully seated.

Figure 1.

PRECAUTION:

ONLY CHARGE BATTERY IN THIS DEVICE OR IN AN APPROVED

CHARGER. (SEE APPROVED ACCESSORIES GUIDE.)

Connect the AC power supply (consult Approved Accessories Guide) by
plugging the round connector into the receptacle on the side of the concen-
trator. See Figure 2. Plug the other end of the AC power supply into a power
outlet. Always use caution when inserting the power supply to a wall outlet.

Charger requirements are universal to support a wide variety of international
markets, so it can be plugged into an outlet with 100-240V AC, 50-60 Hz.
Allow the battery to charge for a minimum of three hours before use. Once
the battery is completely charged, the device can run on battery for approxi-
mately 2-3 hours at a setting of 2.0.

Note: Battery run time may vary based on breathing rate, age of battery, and environmental
conditions. See displayed text on device for battery charge status.

DL-00096 rev 7.indd 2

6/1/2009 4:48:26 PM

Advertising