Alarm Message

Check Vents

Low Battery



XX: Service!*

*Value: 21-50

Description

Device is unable to maintain
oxygen purity.

Estimated battery life less
than 17 minutes.


Service required.

Action

Be sure air inlet/outlet
has not been blocked. If
alarm persists, contact
Customer Service.
Charge the battery pack
by plugging in to power
supply.

Note: The message will be
automatically cleared when
plugged in to power supply.

Contact Customer
Service.

High Priority Alarms

Note: These alarms will allow the Portable Oxygen Concentrator to continue operating.

Alarm Message

Check Cannula





XX: Service*

*Value: 51-70

Description

No breath detected for 5
minutes.




Service required.

Action
Check the cannula con-

nection. Be sure to

breathe through nose. If

alarm persists, contact

Customer Service.

Note: The message will be
automatically cleared when
breathing is detected.

Contact Customer

Service.